您当前的位置:首页 > 资讯中心 > 行业标准 > 正文

Differences between radiant panel test methods ISO-ASTM

2010-5-20 17:31:07 来源:互联网 编辑:未知 阅读:566次 字体:  我要投稿

 

 

ISO9239

ASTME648

Chamber 1400×500and 725 from spccimen  to  top  of  chanber

Chamber 1400×500and 710 from spccimen  to  top  of  chanber

Chanber made from calcium silicate boar 13mm thick, 650kg/m³ density

Chanber made from calcium silicate boar 13mm thick, 740kg/m³ density

Window(110±10)×(1100±10)

Window100×1100

Air access around test specimen holder is 0.23±0.03

Air access around test specimen holder is 0.258--0.3225

Inner edge of chimney is 60±10mm ftom outer edge of chamber

Inner edge of chimney is 51mm ftom outer edge of chamber

Radiabt panel angle 30±1°

Radiabt panel angle 30±5°

0 mark is 100 ±20mm from lowest edge of panel

0 mark is 89±3mm from lowest edge of panel

Panel is (90±10)mm from inner edge of chamber

Panel position relative to chamber wall not specified (but see E 970)

Exhaust hood 30-50mm above chimney (only in drawing)

Exhaust hood >76mm above chimney

Hot wire not specified

Hot wire anemometer specified

Anemometer measurement at 250mm up from bottom of stack

Anemometer measurement at 152mm down from top of stack

Flow with panel off,dummy specimen in place ,door closed=(2.5±0.2)m/s

Flow with panel on,dummy specimen in place , door closed=(1.27±0.25)m/s

Specimen holder-spesimen exposure opening (200±3)×(1015±10)

Specimen holder-spesimen exposure opening 200×1000

Specimen holder - no long studs specified

Specimen holder - 70mm long studs on bottom

Specimen holder external dimension 1140±10 × 325±10

Specimen holder external dimension 1140 × 320

Stack thermocouple at (150±2)mm from top of chimney(same position as ASTM E 648 pyrometer)

No stack thermocouple

Position of stack is not fully defined - Figure 4 suggests center line

 

Pilot burner gas flow (0.026±0.2)1/s=(1.56 ±0.12)1/min(1.44-1.68/min)

Pilot burner gas flow 0.085-0.100lm³/h=1.417-1.6671/min

Pilot burner position;3mm above specimen holder

Pilot burner position;no more than 5°from  horizontal

Pilot burner positioned so that flames generated from the lower line of holes impinge on the specimen (10±2)mm from the zero position.

Pilot burner flame impinges at 0mm mark

Dummy(calibration)board;

(20±1)mm thick

Calcium silicate

(850±100)kg/m³

(250±10)×(1050±20)

(26±1)mm holes at 110,210,...910

Dummy(calibration)board;

19mm thick

Calcium silicate(Fig 7 says inorganic millboard)

740kg/m³(or 580 in Fig 7)

250×1070

27mm holes at 100,200,...900

No bearing plate

Bearing plate to help position HFM (see clause 6.7)

Flux profile procedure:

1)Measure flux at 410 mm point (5.1±0.2kw/)

2)Fluxmeter 2-3 mm above surface

3)Measure flux after 30 s

4)Measure at 110,210,...910 points

5)Criteria on all 9 points

6)Recheck 410 mm position

7)Remove dummy specimen and close door.Wait 5 mins then read chamber temperature and pyrometer

Flux profile procedure:

1)Measure flux at 400 mm point (4.8-5.2kw/)

2)Fluxmeter 1.6-3.2 mm above surface

3)Measure flux after (30±10) s

4)Measure at 100,200,...900 points

5)Criteria on 200,400 and 600 points ONLY

6)Recheck 410 mm position

7)Slide door open to remove dummy specimen. Wait 30 mins then read chamber temperature and pyrometer

Daily Check:

1)No dummy specimen, door closed

2)Black body temp. Within ±5°C

3)Chamber temp. within ±10°C

Daily Check:

1)Door open

2)Black body temp. Within ±5°C

3)Chamber temp. within ±5°C

Specimen:

Dims: (1050±5)×(230±5)

Specimen:

Dims: 1050×250

Test Procedure:

1)Light piout burner at least 50 mm away from zero point

2)Push specimen into chamber (time 0)

3)120s: put pilot is place (flame at 10 mm position)-exposure for 10 mins

4)720s: remove pilot burner and turn off flames

5)1800s: finish test

Test Procedure:

1)Light piout burner at least 50 mm away from zero point

2)Push specimen into chamber (time 0)

3)300s: put pilot is place (flame at 10 mm position)-exposure for 10 mins

4)600s: remove pilot burner and turn off flames

5)Flameout: finish test (or 600s if no pripagation)

Record:

10-,20-and 30- minute flame spread to nearest 10 mm flame spread at flameout time to nearest 10 mm time for flame to reach each rake position maximum flame spread to nearest 10 mm

Record:

maximum flame spread to nearest 1 mm

 

Notes on ASTM E 970

Shows lower edge of panel 105±10mm from inner edge of chamber (ISO 90±10)

Pyrometer calibrated over range 460 - 510°C

Extra calibration hole at 980 mm mark

Test procedure is

1)Light piout burner at least 50 mm away from zero point

2)Push specimen into chamber (time 0)

3)120s: put pilot is place (flame at 10 mm position)-exposure for 10 mins

4)240s: remove pilot burner and turn off flames

5)Flameout: finish test (or 240s if no pripagation)

  您看到此篇文章时的感受是:
0
0
0
0
0
0
0
0
欠扁 支持 很棒 找骂 搞笑 软文 不解 吃惊
已有 0 条评论
网友评论
正在加载数据,请稍等......
我要说两句